High-performance liquid chromatography is an analytical technique that separates components in a liquid sample by passing them through a packed column under pressure. A pump delivers a mobile phase at a controlled flow rate, and an injector introduces the sample into the stream. Differences in how analytes partition between the mobile phase and the stationary phase cause them to exit the column at different times. Detection then records a signal proportional to the amount of each separated substance. The resulting chromatogram provides retention times and peak areas for identification and quantification.
Instrumentation includes a solvent delivery system, an autosampler, a column oven, and one or more detectors. Reversed-phase columns with chemically modified silica are widely used, but normal-phase, ion-exchange, size-exclusion, and affinity modes exist for specific separations. Detectors may rely on ultraviolet absorbance, fluorescence, refractive index, or mass spectrometry. Column temperature, mobile phase composition, and flow rate are adjusted to improve resolution. System pressure is monitored because rising pressure can indicate column blockage or deteriorating packing.
System suitability testing is performed before and during analytical runs to confirm that the instrument and method are working as expected. Typical checks include retention time, peak area precision, resolution between critical pairs, tailing factor, and theoretical plate count. Acceptance criteria are set in the method or pharmacopeial monograph. If a suitability check fails, the run may be rejected and the instrument or sample preparation may need investigation. This practice helps prevent release of data from a system that has drifted out of control.
Quality control samples are inserted at intervals to monitor accuracy and precision throughout a batch. Blank samples detect contamination, while spiked samples assess recovery from the sample matrix. Calibration standards establish the relationship between detector response and concentration, and control samples are prepared independently from them whenever possible. Laboratories also participate in proficiency testing and maintain audit trails, instrument logs, and reagent records. Ongoing review of control charts can reveal trends before they cause out-of-specification results.
Method validation demonstrates that an HPLC procedure is suitable for its intended purpose. Common validation parameters include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantification, and robustness. Accuracy reflects agreement with a reference value, while precision describes repeatability under defined conditions. Specificity shows whether the method can measure the analyte in the presence of impurities or matrix components. Validation documents are reviewed before a method is used for routine testing or regulatory submissions.
| Property | Value | Notes |
|---|---|---|
| Separation principle | Differential partitioning | Analytes distribute between mobile and stationary phases. |
| Mobile phase | Liquid solvent mixture | Composition controls retention and selectivity. |
| Stationary phase | Packed column particles | Often chemically bonded silica. |
| Typical detector | UV-Vis or photodiode array | Mass spectrometry is also common. |
| Common synonym | High-performance liquid chromatography | Abbreviated as HPLC. |
High-performance liquid chromatography, or HPLC, separates dissolved compounds by passing a liquid mobile phase through a packed column. Components distribute differently between the stationary phase and the moving liquid, so they travel at different speeds and exit at different times. A detector records these eluting bands as peaks, and peak area or height relates to amount. The technique supports testing in pharmaceuticals, foods, environmental samples, and industrial chemicals. Quantification usually depends on calibration with known standards.
Several separation modes exist, including reversed-phase, normal-phase, ion-exchange, size-exclusion, and hydrophilic interaction liquid chromatography. Reversed-phase uses a nonpolar stationary phase with a polar mobile phase and is widely applied to small organic molecules. Gradient elution changes mobile phase composition during the run, while isocratic elution keeps it constant. Column chemistry, particle size, temperature, flow rate, and mobile phase pH all influence retention and resolution. Method development selects conditions that separate analytes from matrix components and from each other.
Detection commonly uses ultraviolet-visible absorbance, fluorescence, refractive index, or mass spectrometry. Ultraviolet detection depends on molecular chromophores that absorb light at specific wavelengths. Mass spectrometry provides mass information and sensitive quantification, often after electrospray ionization. Before sample batches, performance checks examine resolution, elution time repeatability, peak symmetry, and plate count. Matrix effects and co-elution remain recognized uncertainties; formal validation studies and orthogonal detection help address them. Detector choice depends on analyte properties and required sensitivity.
Method validation evaluates accuracy, precision, specificity, linearity, range, detection limit, quantitation limit, and robustness. Regulatory guidance for pharmaceuticals, foods, and environmental testing defines expected documentation and acceptance criteria. Verification confirms that a validated method works in a specific laboratory with its own instruments and reagents. Calibration curves use reference standards with known purity and traceability, while measurement uncertainty is estimated from validation data, control charts, and collaborative studies. The scope of validation depends on the method's intended use.
Routine quality control monitors retention time shifts, baseline noise, system pressure, and peak shape. Trends can reveal column aging, mobile phase preparation errors, detector drift, or sample degradation. Corrective actions may include replacing the column, preparing fresh mobile phase, or recalibrating the detector. Stability testing often uses HPLC to measure parent compound loss and degradation product formation. Open questions remain about how accelerated stability results extrapolate to long-term storage under varied conditions.
Quality control for HPLC testing combines scheduled checks, documented procedures, and review of results. Before sample analysis, system suitability testing confirms that the instrument, column, and method meet predefined criteria. Common criteria include resolution between critical peaks, retention time precision, peak tailing, and theoretical plate count. Failure triggers investigation before results are reported. Records link raw data, calculations, instrument logs, and analyst identity to each batch, supporting audits and repeat analysis.
Key performance measures include retention time, peak area, peak height, resolution, tailing factor, and plate count. Retention time helps identify a peak under fixed conditions, but confirmation often requires a second method or detector. Peak area and height relate to concentration through calibration curves, which may be linear or nonlinear depending on the detector response. Resolution describes separation between adjacent peaks, while tailing factor and plate count describe peak shape and column efficiency. Performance checks verify these values before and during a run to confirm that the instrument is performing within limits.
High-performance liquid chromatography testing separates components of a liquid sample by forcing a mobile phase through a packed column. The stationary phase inside the column interacts with analytes to different degrees, so each compound exits at a characteristic retention time. A pump delivers solvent at controlled flow and pressure, while an injector introduces a precise sample volume. Detectors such as ultraviolet-visible, fluorescence, refractive index, or mass spectrometric instruments record the separated bands. The resulting chromatogram provides qualitative and quantitative information about the mixture.
Separation modes differ by the chemistry of the stationary phase and the composition of the mobile phase. Reversed-phase testing uses a nonpolar column and polar solvents, making it common for pharmaceutical, environmental, and food analytes. Normal-phase testing uses a polar column and nonpolar solvents for compounds that are poorly retained in reversed-phase systems. Ion-exchange and ion-pair methods separate charged species, while size-exclusion methods sort molecules by hydrodynamic volume. Gradient elution changes solvent strength over time to resolve complex mixtures, and isocratic elution holds solvent composition constant for simpler assays.
Routine quality control uses system suitability, blank injections, check standards, and control samples to detect drift or contamination. System suitability criteria may specify minimum resolution, maximum tailing factor, and a permitted range for repeated injections. Blank injections reveal carryover or solvent contamination, while check standards confirm calibration accuracy over a batch. Control samples with known analyte levels can show whether results remain within statistical limits. When a control result falls outside limits, the analyst investigates the cause and may invalidate affected results before repeating the batch.
Documentation and traceability are central to regulated HPLC testing. Records typically include instrument logs, column history, mobile-phase preparation, sample preparation, injection sequences, raw chromatograms, and audit trails. Electronic systems may require user access controls, time-stamped changes, and backup procedures. Training records show that analysts are qualified for assigned methods. Audits and inspections check whether written procedures match actual practice and whether deviations are documented. These controls support reproducibility and allow results to be reconstructed if questions arise later.
In some reactions, protons and hydroxide may directly act as acid and base in term of specific acid and specific base catalysis. But more often groups in substrate and active site act as Brønsted–Lowry acid and base. This is called general acid and general base theory. The easiest way to distinguish between them is to check whether the reaction rate is determined by the concentrations of the general acid and base. If the answer is yes then the reaction is the general type. Since most enzymes have an optimum pH of 6 to 7, the amino acids in the side chain usually have a pKa of 4~10. Candidate include aspartate, glutamate, histidine, cysteine. These acids and bases can stabilise the nucleophile or electrophile formed during the catalysis by providing positive and negative charges.
Kunitz-type serine protease inhibitor APEKTx1 is a peptide toxin derived from the sea anemone Anthopleura elegantissima. This toxin has a dual function, acting both as a serine protease inhibitor and as a selective and potent pore blocker of Kv1.1, a shaker related voltage-gated potassium channel. APEKTx1 is a potent toxin purified from the sea anemone A. elegantissima. Besides APEKTx1, other toxins such as APETx1, APE1-1, APE1-2, APE2-2, ApC, and APETx2 have been identified in A. elegantissima. This peptide has 65 amino acids crosslinked by 3 disulphide bridges, and has a molecular mass of 7475 Da. It acts as a monomer. The toxin belongs to the type 2 sea anemone peptides targeting voltage-gated K channels. Other type 2 toxins are the kalicludines from Anemonia sulcata, which selectively block Kv1.2 channels, and SHTX II from Stichodactyla haddoni. Structural homology is also shared with the basic pancreatic trypsin inhibitor (BPTI), a very potent Kunitz-type protease inhibitor, and dendrotoxins (DTX I and α-DTX), which are potent inhibitors of voltage-gated potassium channels.
Affinity chromatography is a method of separating a biomolecule from a mixture, based on a highly specific macromolecular binding interaction between the biomolecule and another substance. The specific type of binding interaction depends on the biomolecule of interest; antigen and antibody, enzyme and substrate, receptor and ligand, or protein and nucleic acid binding interactions are frequently exploited for isolation of various biomolecules. Affinity chromatography is useful for its high selectivity and resolution of separation, compared to other chromatographic methods.
Chattopadhyay's contributions in membrane and receptor biology and biophysics have been recognized by several awards and prizes. These include The World Academy of Sciences (TWAS) Prize, Shanti Swarup Bhatnagar Award, Ranbaxy Research Award, Prof. G.N. Ramachandran Gold Medal, SERB Distinguished Fellowship, Prof. G.N. Ramachandran 60th Birthday Medal and J.C. Bose Fellowship. He is an elected Fellow of The World Academy of Sciences, Royal Society of Biology, Royal Society of Chemistry, and all the Indian Academies of Science. Fellow, The World Academy of Sciences (2017) Fellow, The Royal Society of Biology (2017) Fellow, The Royal Society of Chemistry (2013) Fellow, Indian National Science Academy (2005) Fellow, Indian Academy of Sciences (1999) Fellow, The National Academy of Sciences, India (1998) Fellow, West Bengal Academy of Science & Technology (2010) Fellow, Andhra Pradesh Akademi of Sciences (2003) Fellow, Telangana Academy of Sciences (2015) Google Scholar
Sources: en.wikipedia.org
Greater emphasis on risk management Introduces “Special Requirements” Introduces “Critical Items” Measure: Requirements conformance Measure: Delivery performance Adopt proven product development processes Eliminate “recurring corrective actions” AS9100 Revision C was released in January, 2009, with considerable delay in application of the new version in audits, largely due to the delay in the release of AS9101 Revision D and auditor training to the increased auditing rigor of that update. AS9100 Revision D (2016), Quality Management System – Requirements for Aviation, Space and Defense Organizations The update of AS9100 from revision C to D includes the full text of ISO 9001:2015. In addition to aligning the structure of the aviation, space and defense requirements to the new structure of ISO 9001:2015, the following key changes were implemented:
The prevalence of abdominal obesity is increasing in Western populations, possibly due to a combination of low physical activity and high-calorie diets, and also in developing countries, where it is associated with the urbanization of populations. Other environmental factors, such as maternal smoking, estrogenic compounds in the diet, and endocrine-disrupting chemicals may be important also. Hypercortisolism, such as in Cushing's syndrome, also leads to central obesity. Many prescription drugs, such as dexamethasone and other steroids, can also have side effects resulting in central obesity, especially in the presence of elevated insulin levels. There are various ways of measuring abdominal obesity including: Absolute waist circumference (>102 cm (40 in) in men and >88 cm (35 in) in women) Waist–hip ratio (the circumference of the waist divided by that of the hips of >0.9 for men and >0.85 for women) Waist-stature ratio (waist circumference divided by their height, >0.5 for adults under 40 and >0.6 for adults over 50) Sagittal Abdominal Diameter
the cAMP signal pathway and the phosphatidylinositol signal pathway. When a ligand binds to the GPCR it causes a conformational change in the GPCR, which allows it to act as a guanine nucleotide exchange factor (GEF). The GPCR can then activate an associated G protein by exchanging the GDP bound to the G protein for a GTP. The G protein's α subunit, together with the bound GTP, can then dissociate from the β and γ subunits to further affect intracellular signaling proteins or target functional proteins directly depending on the α subunit type (Gαs, Gαi/o, Gαq/11, Gα12/13). GPCRs are an important drug target, and approximately 34% of all Food and Drug Administration (FDA) approved drugs target 108 members of this family. The global sales volume for these drugs is estimated to be 180 billion US dollars as of 2018. It is estimated that GPCRs are targets for about 50% of drugs currently on the market, mainly due to their involvement in signaling pathways related to many diseases i.e. mental, metabolic including endocrinological disorders, immunological including viral infections, cardiovascular, inflammatory, senses disorders, and cancer. The long ago discovered association between GPCRs and many endogenous and exogenous substances, resulting in e.g. analgesia, is another dynamically developing field of the pharmaceutical research.
Sources: en.wikipedia.org
CGRP antagonists can be used for both acute migraine treatment as well as prophylactically. CGRP is a neuropeptide which is thought to induce migraines via vasodilation of cranial arteries. CGRP can also release inflammatory agents and cause nervous system sensitization. It is theorized that by antagonizing the CGRP receptor of the trigeminal ganglia, lowered CGRP is released and less migraine occurs. Erenumab is a highly selective human monoclonal antibody which is a promising new development in migraine treatment. It has low risk of hepatoxicity like gepants can have, due to being mostly eliminated via proteolysis. There have been some studies suggesting the benefit of using melatonin for prophylaxis of migraine, however, there is a lack of strength of evidence due to a low number of studies as well as conflicting results. Melatonin has a good safety profile but there have been rare instances of serious side effects. More studies are needed in order to suggest the therapeutic use of melatonin for prophylaxis of migraine.
Bone morphogenetic proteins (BMPs) are a group of growth factors also known as cytokines and as metabologens. Professor Marshall Urist and Professor Hari Reddi discovered their ability to induce the formation of bone and cartilage, BMPs are now considered to constitute a group of pivotal morphogenetic signals, orchestrating tissue architecture throughout the body. The important functioning of BMP signals in physiology is emphasized by the multitude of roles for dysregulated BMP signalling in pathological processes. Cancerous disease often involves misregulation of the BMP signalling system. Absence of BMP signalling is, for instance, an important factor in the progression of colon cancer, and conversely, overactivation of BMP signalling following reflux-induced esophagitis provokes Barrett's esophagus and is thus instrumental in the development of esophageal adenocarcinoma. Recombinant human BMPs (rhBMPs) are used in orthopedic applications such as spinal fusions, nonunions, and oral surgery. rhBMP-2 and rhBMP-7 are Food and Drug Administration (FDA)-approved for some uses. rhBMP-2 causes more overgrown bone than any other BMPs and is widely used off-label.
AlphaKnot is a scientific database and web server for detecting, classifying, and visualizing protein knots and other forms of protein-chain entanglement. It was developed to facilitate the analysis of protein structures predicted by AlphaFold and other machine-learning methods, but can also be used to analyze experimentally determined structures. The current version, AlphaKnot 2.0, combines two closely related components: a precomputed database containing proteins identified as knotted in large-scale structure-prediction datasets, and an analysis server that allows users to investigate the topology of individual protein structures in greater detail.
ProIAPP consists of 67 amino acids, which follow a 22 amino acid signal peptide which is rapidly cleaved after translation of the 89 amino acid coding sequence. The human sequence (from N-terminus to C-terminus) is: (MGILKLQVFLIVLSVALNHLKA) TPIESHQVEKR^ KCNTATCATQRLANFLVHSSNNFGAILSSTNVGSNTYG^ KR^ NAVEVLKREPLNYLPL. The signal peptide is removed during translation of the protein and transport into the endoplasmic reticulum. Once inside the endoplasmic reticulum, a disulfide bond is formed between cysteine residues numbers 2 and 7. Later in the secretory pathway, the precursor undergoes additional proteolysis and posttranslational modification (indicated by ^). 11 amino acids are removed from the N-terminus by the enzyme proprotein convertase 2 (PC2) while 16 are removed from the C-terminus of the proIAPP molecule by proprotein convertase 1/3 (PC1/3). At the C-terminus Carboxypeptidase E then removes the terminal lysine and arginine residues. The terminal glycine amino acid that results from this cleavage allows the enzyme peptidylglycine alpha-amidating monooxygenase (PAM) to convert the terminal glycine to an amine group (releasing glycolate). After this step, the transformation from the precursor protein proIAPP to the biologically active IAPP (amylin) is complete (IAPP sequence: KCNTATCATQRLANFLVHSSNNFGAILSSTNVGSNTY-NH2).
Sources: en.wikipedia.org
HPLC separates and detects individual compounds in a liquid sample, producing peaks at characteristic retention times. Peak area or height can be used to estimate concentration when calibrated with known standards. It does not identify unknown compounds with certainty unless additional detectors or reference materials are used.
Pressure drives the liquid mobile phase through a column packed with small particles. Without pressure, flow would be very slow or stop because the packed bed resists liquid movement. Modern pumps maintain a steady flow despite the resistance.
A chromatogram is a plot of detector signal against time after sample injection. Each peak represents a compound or group of compounds eluting from the column. Retention time and peak area are the main measurements read from the plot.
Method validation is the documented process of showing that an HPLC procedure produces reliable results for a defined purpose. It examines parameters such as accuracy, precision, specificity, linearity, and robustness. Regulators and quality systems often require validation before routine use.