This is a working overview of reversed-phase, written for readers who want more than a one-paragraph summary but less than a textbook.
Reviewed 2026-07-18. Anything still debated is marked as such rather than presented as settled.
Routine quality control monitors retention time shifts, baseline noise, system pressure, and peak shape. Trends can reveal column aging, mobile phase preparation errors, detector drift, or sample degradation. Corrective actions may include replacing the column, preparing fresh mobile phase, or recalibrating the detector. Stability testing often uses HPLC to measure parent compound loss and degradation product formation. Open questions remain about how accelerated stability results extrapolate to long-term storage under varied conditions.
Quality control for HPLC testing combines scheduled checks, documented procedures, and review of results. Before sample analysis, system suitability testing confirms that the instrument, column, and method meet predefined criteria. Common criteria include resolution between critical peaks, retention time precision, peak tailing, and theoretical plate count. Failure triggers investigation before results are reported. Records link raw data, calculations, instrument logs, and analyst identity to each batch, supporting audits and repeat analysis.
Method validation evaluates accuracy, precision, specificity, linearity, range, detection limit, quantitation limit, and robustness. Regulatory guidance for pharmaceuticals, foods, and environmental testing defines expected documentation and acceptance criteria. Verification confirms that a validated method works in a specific laboratory with its own instruments and reagents. Calibration curves use reference standards with known purity and traceability, while measurement uncertainty is estimated from validation data, control charts, and collaborative studies. The scope of validation depends on the method's intended use.
Several separation modes exist, including reversed-phase, normal-phase, ion-exchange, size-exclusion, and hydrophilic interaction liquid chromatography. Reversed-phase uses a nonpolar stationary phase with a polar mobile phase and is widely applied to small organic molecules. Gradient elution changes mobile phase composition during the run, while isocratic elution keeps it constant. Column chemistry, particle size, temperature, flow rate, and mobile phase pH all influence retention and resolution. Method development selects conditions that separate analytes from matrix components and from each other.
Detection commonly uses ultraviolet-visible absorbance, fluorescence, refractive index, or mass spectrometry. Ultraviolet detection depends on molecular chromophores that absorb light at specific wavelengths. Mass spectrometry provides mass information and sensitive quantification, often after electrospray ionization. Before sample batches, performance checks examine resolution, elution time repeatability, peak symmetry, and plate count. Matrix effects and co-elution remain recognized uncertainties; formal validation studies and orthogonal detection help address them. Detector choice depends on analyte properties and required sensitivity.
| Property | Value | Notes |
|---|---|---|
| Retention time RSD | ≤1% for five replicate injections | Typical criterion; method-specific limits apply. |
| Resolution | ≥1.5 between critical pair | Baseline separation is generally desired. |
| Tailing factor | ≤2.0 | Measures peak symmetry. |
| Theoretical plates | ≥2000 per column | Method-dependent; higher values indicate greater efficiency. |
| Peak area RSD | ≤2% for replicate injections | Reflects autosampler and detector precision. |
Method validation demonstrates that an HPLC procedure is suitable for its intended purpose. Common validation parameters include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantification, and robustness. Accuracy reflects agreement with a reference value, while precision describes repeatability under defined conditions. Specificity shows whether the method can measure the analyte in the presence of impurities or matrix components. Validation documents are reviewed before a method is used for routine testing or regulatory submissions.
System suitability testing is performed before and during analytical runs to confirm that the instrument and method are working as expected. Typical checks include retention time, peak area precision, resolution between critical pairs, tailing factor, and theoretical plate count. Acceptance criteria are set in the method or pharmacopeial monograph. If a suitability check fails, the run may be rejected and the instrument or sample preparation may need investigation. This practice helps prevent release of data from a system that has drifted out of control.
Instrumentation includes a solvent delivery system, an autosampler, a column oven, and one or more detectors. Reversed-phase columns with chemically modified silica are widely used, but normal-phase, ion-exchange, size-exclusion, and affinity modes exist for specific separations. Detectors may rely on ultraviolet absorbance, fluorescence, refractive index, or mass spectrometry. Column temperature, mobile phase composition, and flow rate are adjusted to improve resolution. System pressure is monitored because rising pressure can indicate column blockage or deteriorating packing.
Separation performance depends on particle size, pore size, column length, and the chemistry of the stationary phase. Smaller particles generally improve efficiency but require higher pressure and suitable instrumentation. The mobile phase often contains buffers and organic solvents that influence retention and selectivity. Testing labs select conditions based on the analytes, sample matrix, and required sensitivity. Method development frequently involves screening several columns and solvent mixtures before a final set of conditions is chosen.
High-performance liquid chromatography is an analytical technique that separates components in a liquid sample by passing them through a packed column under pressure. A pump delivers a mobile phase at a controlled flow rate, and an injector introduces the sample into the stream. Differences in how analytes partition between the mobile phase and the stationary phase cause them to exit the column at different times. Detection then records a signal proportional to the amount of each separated substance. The resulting chromatogram provides retention times and peak areas for identification and quantification.
14-3-3 proteins are a family of conserved regulatory molecules that are expressed in all eukaryotic cells. 14-3-3 proteins have the ability to bind a multitude of functionally diverse signaling proteins, including kinases, phosphatases, and transmembrane receptors. More than 200 signaling proteins have been reported as 14-3-3 ligands. Elevated amounts of 14-3-3 protein in cerebrospinal fluid are usually a sign of rapid neurodegeneration; a common indicator of Creutzfeldt–Jakob disease.
Back pressure is the term used for the hydraulic pressure required to create a flow through a chromatography column in high-performance liquid chromatography, the term deriving from the fact that it is generated by the resistance of the column, and exerts its influence backwards on the pump that must supply the flow. Back-pressure is a useful diagnostic feature of problems with the chromatography column. Rapid chromatography is favoured by columns packed with very small particles, which create high back-pressures. Column designers use "kinetic plots" to show the performance of a column at a constant back-pressure, usually selected as the maximum that a system's pump can reliably produce. Exhaust pulse pressure charging Expansion chamber Scalar quantity
In order to regulate the secretion of ACTH, many substances secreted within this axis exhibit slow/intermediate and fast feedback-loop activity. Glucocorticoids secreted from the adrenal cortex work to inhibit CRH secretion by the hypothalamus, which in turn decreases anterior pituitary secretion of ACTH. Glucocorticoids may also inhibit the rates of POMC gene transcription and peptide synthesis. The latter is an example of a slow feedback loop, which works on the order of hours to days, whereas the former works on the order of minutes. The half-life of ACTH in human blood is reported to be between ten and 30 minutes. ACTH consists of 39 amino acids, the first 13 of which (counting from the N-terminus) may be cleaved to form α-melanocyte-stimulating hormones (α-MSH) (this common structure is responsible for excessively tanned skin in Addison's disease). After a short period of time, ACTH is cleaved into α-melanocyte-stimulating hormone (α-MSH) and CLIP, a peptide with unknown activity in humans. In the human body, total weight ACTH is 4540 Da.
The Bailey peptide synthesis is a name reaction in organic chemistry developed 1949 by J. L. Bailey. It is a method for the synthesis of a peptide from α-amino acid-N-carboxylic acid anhydrides (NCAs) and amino acids or peptide esters. The reaction is characterized by short reaction times and a high yield of the target peptide. The reaction can be carried out at low temperatures in organic solvents. The residues R1 and R2 can be organic groups or hydrogen atoms, R3 is the used amino acid or peptide ester: The reaction mechanism is not known in detail. Supposedly, the reaction begins with a nucleophilic attack of the amino group on the carbonyl carbon of the anhydride group of the N-carboxylic acid anhydride (1). After an intramolecular proton migration, a 1,4-proton shift and the cleavage of carbon dioxide follows, resulting in the peptide bond in the final product (2):
Sources: en.wikipedia.org
In 80–85% of cases, the ALK detected in ALK-positive ALCL is a NPM1-ALK fusion protein. It is made by a fusion of NPM1 gene, which makes nucleophosmin 1, located on the long or "q" arm of chromosome 5 at position 35 (notated as 5q35) with the ALK gene located on the short or "p" arm of chromosome 2 at position 23 (notated as 2p23) to form a chimeric gene notated as (2;5)(p23;q35). In 13% of cases ALK fuses with the TPM3 gene or in <1% of cases for each of the following genes: TFG, ATIC, CLTC, TPM4, MSN, RNF213 (also termed ALO17), MYH9, or TRAF1. All of these fusion proteins are considered to act like NPMI-ALK in possessing high ALK activity that promotes the development and progression ALK-positive ALCL by activating the cell signaling pathways cited in the Introduction. 15% Of individuals with ALK-positive ALCL also have point mutations in the NOTCH1 gene. While most of these abnormalities are thought to be detrimental not all are. For example, DUSP22 gene rearrangements are associated with favorable outcomes in ALK-positive (as well as ALK-negative) ALCL.
Lab Matters: Each issue of APHL's quarterly magazine, Lab Matters, considers a pressing health issue of importance to public health laboratories. Bridges: Connecting the nation's environmental laboratories, Bridges is APHL's semi-annual newsletter for environmental health professionals. Social Media: APHL is active on Twitter, Facebook and LinkedIn, and has video channels on both YouTube and Vimeo. Blog: APHL's Public Health LabLog features recent and interesting stories about the work of APHL and public health laboratories. 1899: American Public Health Association formed the Committee of Laboratories 1921: Southern Public Health Laboratory Association (SPHLA) formed 1927: SPHLA became State Laboratory Directors Conference and opened membership to other states 1939: SPHLA changed its name to Conference of State and Provincial Laboratory Directors 1951: Association of State and Territorial Public Health Laboratory Directors founded 1998: Renamed the Association of Public Health Laboratories (APHL) and became a more inclusive organization with new membership categories
Projected to go undrafted by Sports Illustrated, who criticized his lack of "balance, body control and ability in space," Kuper was ranked as the No. 28 offensive guard available in the 2006 NFL draft. Kuper was eventually selected in the fifth round, 161st overall, by the Denver Broncos. He was the highest selected North Dakota lineman since Todd Thomas in 1981. Kuper started 79-of-90 games played (68 at right guard) while playing his entire eight-year NFL career with the Broncos. On June 4, 2010, the Broncos announced that they had signed Kuper to a five-year deal worth $25.5 million. The new contract made him the second-highest-paid offensive lineman in Broncos' history, second only to former left tackle Matt Lepsis. In 2010, the Denver Broncos announced that Kuper had been selected as co-captain of the Broncos offense, along with Kyle Orton and Daniel Graham. During the regular season finale in 2011 against the Kansas City Chiefs, Kuper suffered a gruesome broken ankle that ended his season. That season Kuper made the Pro Bowl, but withdrew due to the injury and needed surgery. A three-time captain, Kuper was recipient of the Ed Block Courage Award in 2012 as voted on by his teammates. Kuper retired on March 11, 2014.
Sources: en.wikipedia.org
System suitability is typically performed before each batch or according to the validated method and laboratory procedure. Some long runs include periodic checks during analysis. The required frequency depends on regulatory expectations and method performance.
Retention time drift can result from changes in mobile phase composition, column temperature, pump flow, or column age. A gradual shift often points to column degradation. A sudden shift may indicate a leak, mixing error, or incorrect mobile phase.
Retention time alone cannot confirm identity because different compounds may elute at similar times. Coupling HPLC with mass spectrometry or comparing against authenticated standards increases confidence. Confirmation usually requires orthogonal data.
HPLC separates and quantifies compounds in a liquid sample. Detectors produce a response proportional to the amount of a compound passing through the flow cell. Identification by retention time requires comparison with a known standard.